1376
Trigge r
Trig Delay
Controls a time delay during which the analyzer will wait to begin a sweep after meeting the trigger
criteria. You can use negative delay to pre-trigger the instrument in time domain or FFT.
Q (Demodulated)
Pressing this key, when it is not selected, selects the main receiver's output Q voltage as the trigger. The
Q (Demodulated) trigger condition is met when the Q voltage crosses the Q voltage trigger level.
Trigger Level
Sets a level for the Q (Demodulated) trigger. When the signal crosses this level, with the chosen slope,
the trigger occurs. If the specific Measurement displays the signal from the chosen sampling point a
Remote Command: :TRIGger[:SEQuence]:IDEMod:SLOPe POSitive | NEGative
:TRIGger[:SEQuence]:IDEMod:SLOPe?
Example: TRIG:IDEM:SLOP POS
Preset: POSitive
State Saved: Saved in instrument state
Initial S/W Revision: Prior to A.02.00
Key Path: Trigger , Baseband I/Q, I (Demodul ated )
Remote Command: :TRIGger[:SEQuence]:IDEMod:DELay <time>
:TRIGger[:SEQuence]:IDEMod:DELay?
:TRIGger[:SEQuence]:IDEMod:DELay:STATe OFF|ON|0|1
:TRIGger[:SEQuence]:IDEMod:DELay:STATe?
Example: TRIG:IDEM:DEL 10 ms
TRIG:IDEM:DEL:STAT ON
Preset: 1 us
OFF
State Saved: Saved in instrument state
Range: –2.5 s to +10 s
Initial S/W Revision: Prior to A.02.00
Key Path: Trigger, Baseband I/Q
Example: TRIG:<meas>:SOUR QDEM
Readback Text: Q
Initial S/W Revision: Prior to A.02.00