884
AMPTD Y Scale
Off
Turns the internal preamp offLow BandSets the internal preamp to use only the low band.The frequency range of the installed (optional) low-band preamp is displayed in square brackets on the
Low Band key label.
Preset: OFF
State Saved: Saved in instrument state
Initial S/W Revision: Prior to A.02.00
Key Path: AMPTD Y Scale, Internal Preamp
Scope: Meas Global
Remote Command: [:SENSe]:POWer[:RF]:GAIN:BAND LOW|FULL
[:SENSe]:POWer[:RF]:GAIN:BAND?
Dependencies: Preamp is not available on all hardware platforms. If the preamp is not present
or is unlicensed, the key is not shown.
If a POW:GAIN:BAND FULL command is sent when a low band preamp is
available, the preamp band parameter is to LOW instead of FULL, and an
"Option not installed" message is generated.
Preset: LOW
State Saved: Saved in instrument state
Initial S/W Revision: Prior to A.02.00
Key Path: AMPTD Y Scale, Internal Preamp
Example: :POW:GAIN OFF
Readback: Off
Initial S/W Revision: Prior to A.02.00
Key Path: AMPTD Y Scale, Internal Preamp
Example: :POW:GAIN ON
:POW:GAIN:BAND LOW
Readback: Low Band
Initial S/W Revision: Prior to A.02.00