1254
Source (Internal)
setup, the value entered run through several levels of validation.
Step Configuration of Step Duration parameter list (Remote Command Only)
This SCPI command is to configure “Step Duration” parameter array of the whole List Sequencer at one time. The
number of array is same as step number defined in “Number of Steps” on page 1221 Number of Steps. As a step is
setup, the value entered run through several levels of validation.
Remote Command: :SOURce:LIST:SETup:WAVeform <string>, <string>,
<string>, …
:SOURce:LIST:SETup:WAVeform?
Example: SOUR:LIST:SET:WAV “CW”,”Off”,”CONT”
SOUR:LIST:SET:WAV?
Notes: The command is to setup below parameter array of whole list sequence.
Wave form <string> - specifies the waveform for playback during the step.
The step can output either a new ARB waveform, continue playback of the
previous waveform, or output a CW tone. The options for specifying these
are:
<filename> - plays the specified waveform from the start. The filename value
is the name of the file within ARB playback memory, it is does not include the
windows path to the file on the HDD. If you enter a filename for a waveform
that does not reside within ARB playback memory, an error is generated.
CONT – continues playback of the ARB file from the previous step
CW – outputs a CW tone
OFF – disable RF output
If input parameter number exceeds the step number defined by Number of
Steps then generate error ", and only those parametes whose index number
falls in number of steps will be updated.
Dependencies: The range is 1 to 1000 which is determined by the number of steps you have
configured. For details see “Number of Steps” on page 1221.
Range: ”filename” | “CW” | “Off” | “CONT”
Initial S/W Revision: A.09.40
Remote Command: :SOURce:LIST:SETup:DURation:TYPE <enum>, <enum>, <enum>,
:SOURce:LIST:SETup:DURation:TYPE?
Example: SOUR:LIST:SET:DUR:TYPE COUN,TIME,CONT
SOUR:LIST:SET:DUR:TYPE?