SET MEASURE TIME | Timing Command Dictionary |
|
|
Figure A-9 shows when the output pin measures occur based on this example.
0NS |
| Measure times (for test cycle 1) |
| 150NS | |||||
|
|
|
|
| |||||
|
| 65NS |
|
|
| ||||
| 15NS | 38NS | 135NS |
| |||||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
cycle starts | cycle |
| ends |
| |||
|
|
|
Figure A-9. SET MEASURE Timing Example
You could use these measure times in conjunction with the force times specified in the SET FORCE TIME example. In the first 150ns test cycle, the measure times occur at 15, 38, 65, and 135ns. In the second test cycle, the measure times occur at 150 plus the specified offsets; that is, 165, 188, 215, and 285ns.
Related Commands |
|
SET END_MEASURE_CYCLE | SET PROCEDURE FILE |
TIME |
|
SET SPLIT_BIDI_CYCLE TIME |
|
FastScan and FlexTest Reference Manual, V8.6_4 |